QRS Universal 12 & 6 Lead ECG
The Universal ECG features accurate results, real-time graphical display and narrative interpretation. The ECG uses the Louvaine algorithm that has the best total accuracy of the 9 popular algorithms compared by cardiologists to a standardized database of ECG tracings. Available as 12-Lead Cable (10 Patient Leads) or 6 Lead Cable.
Features
- Acquire, store and analyze up to 12 channels of high-resolution ECG data
- Print full page reports on standard plain paper in portrait, landscape or A4.
- Review and zoom into data with electronic calipers, enter comments and modify the interpretation before confirming the results.
- Instantly create PDFs or JPEGs of the final report.
- Seamlessly integrate ECG data into your EMR via XML, HL7 and other formats.
- Also available in a 6-channel non-interpretive version.
Technical Specifications
Hub Weight | 280 – 300 grams (0.62 – 0.66 lb) depending on cable options |
Hub Dimensions | 85mm x 91mm x 20mm (3.3” x 3.6” x 0.8”) |
Patient Leads Length | 1 meter (3.3 ft) |
Patient Leads | 6 Lead Cable (4 patient leads) 12 Lead Cable (10 patient leads) |
Case Material | ABS Plastic |
Electrode Connections | 4 mm Banana plug with “tab” or “snap” connectors |
Electrode Labeling | Abbreviations and colors to comply with either IEC or AAMI standards |
Gain/Sensitivity | 5, 10, 20 mm/mV |
Input Range | ±6mV |
Acquisition sample rate | 1000 samples per second (compressed to 500Hz with peak picking and averaging algorithm) |
Frequency Response | 0.05 to 175Hz ±3dB |
Common Mode Rejection | -60dB (minimum) |
Safety Standards | Complies with AAMI EC11, EN60601-1, EN60601-1-2, and EN60601-2-25 |
Supply Voltage | 4 – 16V DC |
Supply Current | <17mA DC |
Permanent Filters | High Pass: 0.05Hz 1st order Low Pass: 170Hz 1st order Baseline Wander: Baseline reset by adaptive zeroing algorithm |
Notch filter (Mains NoiseRejection) | 50Hz 4th order Butterworth, 49.1Hz – 50.9Hz, 60Hz 4th order Butterworth, 59.1Hz – 60.9Hz |
Low pass (Muscle Artifact Filter) | 35Hz 4th order |
Report Capabilities | User selectable Report formats |
Environmental Conditions | Operating Temperature: 0 to 40° C (32 to 104° F) Storage Temperature: -20 to 70° C (-4 to 158° F) Humidity < 85% (non-condensing) |